Pular para o conteúdo principal
GE Healthcare · MR

OrthoMRI Automatic Coil ID Fails.

Symptom

OrthoMRI Automatic Coil ID Fails. Ortho MRI The customer cannot scan because the coil the type of coil being used is not recognized by the system. OrthoMRI The customer cannot scan because the coil the type of coil being used is not recognized by the system.

Error message

OrthoMRI Unknown Coil Error The system does not recognize the coil that is plugged in for use. The high failure part is the coil cable. If this error occurs on more than one coil, the mechanical swing arm cables is the next high failure part.

Applies to

Product
E-Scan XQ. E-Scan XQ.
Software
OrthoMRI 8.2b software Revision. OrthoMRI 8.2b software Revision.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required