Unable to Graphically Prescribe off of a 3-plane Localizer, the Graphic Prescription Window Does not Come Up.
Symptom
Unable to Graphically Prescribe off of a 3-plane Localizer, the Graphic Prescription Window Does not Come Up One of the windows / panes is missing for 3 plane-loc. See long description. The tech scans a 3 plane localizes and then wants to graphically prescribe off of that, but the graphic prescription window does not come up, an error comes up: "no valid localizer exists for current prescription".
Error message
1031 Stored Procedure Failed. Identifyimage failed Error: 1031 Host: aicmr Process: dbwserver File: batch.ec Line: 917 stored procedure failed : -211 - Identifyimage failed 9065 Internal Error, Fork not Allowed NO FREE PID 9065 Host: aicmr Process: dbwserver File: proc.c Line: 187 Internal error, fork not allowed NO FREE PID
Applies to
- Models
- Signa, Excite, Discovery
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.