GE Healthcare · CTLightSpeed26011821410132
LightSpeed 5.X (80/100kW)
Symptom
Scanner stops with Spit Ratio and/or Rotor errors. Long Description for Feedback Only. Scanner stops with Spit Ratio and/or Rotor errors.
Error message
Error: 260118214 (60-0317) Spit Ratio error See Long Description for complete error. Fri Dec 16 23:26:55 2005 Error: 260118214 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Fri Dec 16 23:26:54 2005 Host : Jedi Ermes # : 260118214 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Axial Scan: 10132/2/1 (60-0317)Spit Ratio error. PPC board has detected the rate of tube spits during exposure has reached the limit (see theory of operation, software section).^C Z6,Z7= 4h,0h
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- D3634T (aka 2378211-2 ) Varian PERFORMIX PRO TUBE used on : HP80 AND LS RT4 See Long Description for details Part #: D3634T Part Type: GP TUBES Part Weight: 176.37 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PERFORMIX PRO (VARIAN) TUBE UNIT FRU: Yes Status: Orderable Comments: FLS 11/19/04 (Shawn Fox 212001261) PLEASE remember to complete the Tube report for the tube you are replacing and return it with old tube. The blank Tube report comes with the new tube. If you have any questions, please contact Yolanda Scheper 262-548-2864. If you do not return tube report with old tube, warranty credit can be missed or delayed for your Zone
- Area
- CT
Full cause & step-by-step resolution