LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
CTx00066 Failed shorted DAS/Detector channel detected by PDS See Long Description for details and customer communication. CTx00066 Failed shorted DAS/Detector channel detected by PDS Contact customer and inform them of the PDS detected error regarding the Das/Detector and stress the importance of scheduling time with the Field Engineer to correct, before image quality is compromised. Per the PDS workflow do not close the FE PDS RFS until the issue is resolved. This does not impact the RFS - EMG metric. This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service.
Error message
Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA EXCLUSION CRITERIA *. Error: 230007004 can be used for any row, channel and module that is failing. Error: 230007004 Channel ###, Module ## Short to ground* FAILED X- AXIS EXCLUSION CRITERIA Examples: Sat Sep 17 08:08:02 2005 Error: 230007004 Host: amhvct Process: badchanmgr File: BadChannelMap.cxx Line: 584 Exclusion Test - Row 33:1B, Channel 900, Module 57 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *. Tue Aug 29 05:24:31 2006 Error: 230007004 Host: ct99 Process: badchanmgr File: BadChannelMap.cxx Line: 667 Exclusion Test - Row 1:32A, Channel 418, Module 27 Short to ground* FAILED X-AXIS EXCLUSION CRITERIA *. Error: 230007014 Uncorrectable detector channels have been found for scan See LD for an example of entire error message. Fri Oct 28 09:32:03 2005 Error: 230007014 Host: presbyplano Process: badchanmgr File: badchanmgr.cxx Line: 1194 Uncorrectable detector channels have been found for scan 50029/2/2
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5143018 VCT HALO DIFB (Digital Interface Board) See long description for details If no stock of 5143018 Can use P/N 5194006-3 on Halo Detector Part #: 5143018 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: HALO DIFB PWA FRU: Yes Status: Orderable Comments: ** FE ASSIST ** If no stock of 5143018 Can use P/N 5194006-3 on Halo Detector (11/3/09 212001261) Activated per Expert Case/FEMC Case #6717276 212008958 10-25-07.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.