Pular para o conteúdo principal
GE Healthcare · MRSignaDiscoveryOptima2224980221852722273372224945

CWA ID75856 Can Not Scan.

Symptom

Can Not Scan

Error message

2224980 Host to HPC Interface Error. The HPC received a start scan packet during scan prescription or when it was already scanning Wed Mar 9 10:30:46 2005 Error: 2224980 Host: mrs1-oc0 Process: NSP File: HPC_PORT_ID:hpc_prep_scan Line: 650 Host to HPC Interface Error. The HPC received a start scan packet during scan prescription or when it was already scanning. HPC scan state: 3

2218527 Scanning has stopped. A PSD error occurred. Fri Jul 7 10:38:40 2006 Error: 2227337 Host: TWM1 Process: NSP HPC_PORT_ID:psd scan entry point Line: 0 Scan aborted. Resp triggering failed due to no resp signal present at Systems Cabinet. Fri Jul 7 10:38:40 2006 Error: 2218527 Host: TWM1 Process: scn File: MsaScanOpsScrn.m Line: 909 Scanning has stopped. A PSD error occurred.

2224945 Internal Software Error PSD in infinite loop The IPG watchdog has timed out Tue Nov 15 09:45:10 2005 Error: 2224945 Host: lx-mr Process: NSP File: IPG WDOG:hpc wdog task Line: 15 Internal Software Error. PSD in infinite loop. The IPG watchdog has timed out, PCT was re-started to attempt to recover, but a CV download may be needed to continue.

2218527 Scanning has stopped. Host to HPC Interface Error. The HPC received a s

Applies to

Models
Signa, Discovery, Optima
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues