GE Healthcare · CTLightSpeed
System will not boot-up to application - Getting pink box Warning message "StartMon process is already running.
Symptom
System will not boot-up to application - Getting pink box Warning message "StartMon process is already running. Request rejected." See long description for details System will not boot-up to application - Getting pink box Warning message "StartMon process is already running. Request rejected.
Error message
startMon already running seen in SYSLOG (or messages) See the Long Description for the full error and command to find error startup: startMon: startMon already running Type cd /usr/adm (or /var/log) Type more SYSLOG (or messages) Look for current time and date /Dec 22 Dec 22 07:06:14 5B:ampct startup: startMon: startMon already running
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.