GE Healthcare · CTLightSpeed23000650826390
LightSpeed 4.X - Octane
Symptom
Unable to recon images - document in RFS - critical issue Unable to recon images
Error message
Error: 230006508 RECON JOB FAILED RETRY See the long description for the full error code Thu Jun 30 23:02:44 2005 Error: 230006508 Host: ip01 sbc Process: reconmgr File: rm process job.c Line: 574 Function: Image Generation : SBC Processing Job Type: Prospective Axial Scan: 26390/2/2 Image: 26390/2/3 Recon Seq Id: 15 Reconmgr received job status error RECON JOB FAILED RETRY from IG Control in the ig done event.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405I.2_H4.0M5 for LightSpeed 4.X (LightSpeed 16) with H16 Console (GOC1)(2341923, 2341923-2) - showprods returns 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395173 1204I.2_ H4.0M5 CLASSM LightSpeed/ HiSpeed QX/I Class M Software 2395176 2395170, 2395171, 2399867 LightSpeed 16 Applications Software (Disk 1, Disk 2 and Patch A) 405I.2_H4.0M5 FW47-03 LS2002M5. IRIX only. Class C/M Common w/ LS 2X, 3X, & 4X.
- Area
- CT
Full cause & step-by-step resolution