Service Note 43098 r0 Sep/07 Image Offset in Optima Images
Symptom
Service Note 43098 r0 Sep/07 Image Offset in Optima Images Service Note 43098_0 Sep/07 Image Offset in Optima Images Service Note 43098 _0 Sep/07 Image Offset in Optima Images There have been several occurrences of image shifts or offsets on Optima particularly during a tomo acquisition. The root cause has been narrowed down to either of two possibilities. Both the Buffer boards, 9BUF0104–C, as well as the Coax Connector Board, 5PWB0090–A, were found to produce this effect. They are both located, one per detector, within the detector housing. See Resolution 138549 for Notes
Applies to
- Models
- Optima
- Product
- Optima Optima / Optima NX Optima / Optima NX
- Component
- *Detector (Imaging Detector) *Detector (Imaging Detector) *Detector (Imaging Detector)
- Area
- Nuclear Medicine
Full cause & step-by-step resolution