GE Healthcare · MRSigna

Cannot Archive Cannot archive.

MOD drive failure.

Symptom

Cannot Archive Cannot archive.

Error message

checksense SONYSMO F551 Unexpected DSRT status from the SCSI driver Check MOD log file: cd ~sdc/logfiles;tail -50 arslog|more The error: checksense: SONYSMO-F551 Unexpected DSRT status from the SCSI driver : please report status=0x1 - 'no unit responded' SCSI CMD : 2800012350010000000 Fri May 26 10:51:01 2006 WriteDicomImage: DFopenNext failed to open next file Fri May 26 10:51:01 2006 aw_dicom_store: Failed to write Check MOD log file: cd ~/sdc/logfiles;tail -50 arslog|more The error: checksense: SONYSMO-F551 Unexpected DSRT status from the SCSI driver : please report status=0x1 - 'no unit responded' SCSI CMD : 2800012350010000000 Fri May 26 10:51:01 2006 WriteDicomImage: DFopenNext failed to open next file Fri May 26 10:51:01 2006 aw_dicom_store: Failed to write

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2318598 Sony MOD Drive Model SMO-F551-SD Sony MOD Drive 2318598 (Model #SMO-F551-SD) Sony MOD Drive 2318598 (Model #SMO-F551-SD)
Area
MR

Full cause & step-by-step resolution