P500D System does not shut down fully.
Precision 500D System does not shutdown fully – Symptom 1
Symptom
P500D System does not shut down fully. On opening the system cabinet you hear the clicking of a relay, and the top most LED on the Lime Board "UPS OFF CMD" is blinking. This means the Corona is trying to turn off the UPS, and the UPS outputs are not turning off. Look at the UPS. In this case, the UPS LED's will stay on forever.