Pular para o conteúdo principal
GE Healthcare · MRExcite2249150

Shorted Diode in MRI Devices 1.5T T/R Knee/Foot Array

Symptom

Cannot Scan with Head T/R Driver Error. The system failed to scan either head or body scans. The SSM light indicated T/R Fault.

Error message

2249150 T/R driver fault, head path short. Possibly a shorted diode in the head T/R switch network, or a short circuit load.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2293674-2 MRI KNEE coil for GE 101759 T/R Knee Foot Array
Area
MR

Full cause & step-by-step resolution