XRd2X Systems shipped prior to June 2005 had wrong length Systems Cabinet to Wallstand MIS Cables.

Revolution XRd2X (Lightning) Service Installation Mechanical

Symptom

XRd2X Systems shipped prior to June 2005 had wrong length Systems Cabinet to Wallstand MIS Cables. Standard Length MIS kit (C1602MB) has MIS 11644A, MIS 11755A, MIS 11756A, MIS 11757A and MIS 11759A too long (should be 49 ft, but 65 ft in kit ) Long Length MIS kit (C1602MC) has MIS 11644A, MIS 11755A, MIS 11756A, MIS 11757A and MIS 11759A too short (should be 65 ft, but 49 ft in kit ).