Pular para o conteúdo principal
GE Healthcare · MRSigna

Passwords Expired due to Dead (Exhausted) Battery on the TOD (Time of Day) Module on OCTANE II Computer's System Module.

Symptom

Unable To Login with Passwords Expired Message System asked to change new passwords since old password have been expired.

Error message

WARNING: lost battery backup clock Syslog Error message. Dec 31 19:01:24 4A:ct99 unix: WARNING: lost battery backup clock

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
5245022 RTC (Real Time Clock) module. Dallas chip DS1687-5 NVRAM See Long Description 5245022 RTC (Real Time Clock) module. Per FEMC Part Comments, dated June 30, 2009: This part valid for all MR Octane computer models.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required