GE Healthcare · MRSigna
Passwords Expired due to Dead (Exhausted) Battery on the TOD (Time of Day) Module on OCTANE II Computer's System Module.
Symptom
Unable To Login with Passwords Expired Message System asked to change new passwords since old password have been expired.
Error message
WARNING: lost battery backup clock Syslog Error message. Dec 31 19:01:24 4A:ct99 unix: WARNING: lost battery backup clock
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 5245022 RTC (Real Time Clock) module. Dallas chip DS1687-5 NVRAM See Long Description 5245022 RTC (Real Time Clock) module. Per FEMC Part Comments, dated June 30, 2009: This part valid for all MR Octane computer models.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.