GE Healthcare · MRExcite2218682

Geometry errors Due To Software Update ServicePack-02.

Symptom

Cannot Scan with Geometry Errors Error occurred during Geometry Calculations. Re-prescribe slices/slab in Graphic Rx Screen.

Error message

2218682 Error occurred during Geometry Calculations Fri Jan 19 10:56:01 2007 Error: 2218682 Host: MR1 Process: scn File: RxManagerIF.m Line: 918 . Re-prescribe slices/slab in Graphic Rx Screen.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution