Pular para o conteúdo principal
GE Healthcare · MRExcite22520551121724332

IT-Host Card Driver Initialization Error Due to Fiber Optic Cable.

Symptom

Can Not Scan

Error message

2252055 Diagnostic: IT Host->MGD Hammer. IT-Host card driver initialization error. 2252055 lx-mr diagCoord bit3_diags.c 469 Diagnostic: IT Host->MGD Hammer Iteration: 1 of 1 Action: 1 Task ID: 0x0 Time: 1121724332 Object: Host Diag Executive IT-Host card driver initialization error. This error usually means that the IT-Cable is disconnected or that the cable is damaged.->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution