GE Healthcare · CTLightSpeed18220710494706366543018221120525
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Cradle will move a few inches and stop. Cradle will move in a few inches and stop
Error message
Error: 182207 HW error: Alignment error between Encoder and Potentiometer See Long Description for complete error Tue Aug 5 15:05:39 1997 Error: 182207 Host: ETC Proc:Cradle File: cmcCorrErr.c 1.22 3/11/92 Line: 51 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:39:320 Log Series: 28 HW error: Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm) Error: 182211 Velocity error detected by hardware (LM628 excessive position error). See the Long Description for the full error Tue Aug 29 08:09:30 2006 Error: 182211 Host: ETC Process: Cradle File: cmcLm628IsrTsk.c 1.19 10/16/9 Line: 185 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 20525/2/1 Type: Helical Exception Level: Pri/Soft Time: 08:09:30:260 Log Series: 595 HW error: Velocity error detected by hardware (LM628 excessive position error). Expected Velocity 75000 (.001 mm/sec) Actual Velocity 0 (.01 mm/sec) Pos Err Tol 2540 (.01 mm) Cur Pos 90852 (.01 mm) Cradle Dir: 1 (0 = Out 1 = In [Gantry]).
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required