Pular para o conteúdo principal
GE Healthcare · MRExcite22522392252133

SCP did not respond to the start packet of WavRec.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2252239 Startup Timeout was detected on the Drivers Control Board

2252133 SDO Communication Failure was detected on the CANopen Protocol Node Detecting Failure = SCP Wed Aug 25 08:04:47 2004 2252133 2 MR-OC0 NSP SCP:sdo_segment_upload 1408 SDO Communication Failure was detected on the CANopen Protocol. Node Detecting Failure = SCP. Node ID = 4. Index = 0x3003. SubIndex = 0x0. Error Code = 0x1002.->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR