GE Healthcare · MRExcite2261176
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Faulty SRI 3 assembly.
Symptom
Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis
Error message
2261176 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error:An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5105270 Scan Room Interface 3 Assembly - (SRI 3)
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.