Scanner boots up to applications and shuts down: Replace SBC HSD and load SBC software.
Symptom
System will not boot to or stay in applications. System will not boot or stay in applications
Applies to
- Models
- HiSpeed
- Product
- HSA RP 2.X HiSpeed Advantage HiSpeed Advantage RP 2.X HiSpeed Advantage RP 2.X FMI 25247 HSA/RP 2.5 Y2K Upgrade Special Tools Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2119401 HSArp High Speed Disk in the SRU Cabinet (top most disk) 2.1 GB disk See LD for details Part #: 2119401 Part Type: GP REPAIRABLES Part Weight: 2.27 LBRRepairable: YesReturnable: YesHazardous Material: NDescription: 3.5" 2147MB ST32550N UL, CSA,VDEFRU: YesStatus: OrderableFLS - T3202FW (ST32550N SEAGATE) IS A 2.1 GB DRIVE. USED ON HISPEEDSCAN DATA DISK.
- Area
- CT
Full cause & step-by-step resolution