Pular para o conteúdo principal
GE Healthcare · MR

OrthoMRI Foot/Ankle cable coil is broken OrthoMRI System is not recognizing the coil OrthoMRI Replaced the cable on the…

OrthoMRI Foot/Ankle cable coil is broken OrthoMRI System is not recognizing the coil OrthoMRI Replaced the cable on the Foot/Ankle Coil OrthoMRI EscanXQ OrthoMRI Unknown coil

Symptom

OrthoMRI System is not recognizing the coil.

Error message

OrthoMRI Unknown coil error message.

Applies to

Product
E-Scan XQ. E-Scan XQ.
Software
OrthoMRI 9.1b revision software.
Component
OrthoMRI P/N LNR8800031002 CABLE FOR FOOT/ANKLE COIL. OrthoMRI P/N LNR8800031002 CABLE FOR FOOT/ANKLE COIL. There is a PSDB solution available for deselecting the automatic coil recognition for ALL OrthoMRI systems. If need be, follow that resolution. Then the system can be used as normal once the operator manually selects which coil is being used while on the patient set up page. The most common failure is the particular coil's cable. If multiple coils have this issue the fix could be the mechanical swing arm.
Area
MR

Full cause & step-by-step resolution