GE Healthcare · CTLightSpeed2300150135120052300150352300150241127205412
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Fastcals failed to complete. Fastcals failed to complete
Error message
Error: 230015013 Failure opening a file for reading in the Cal Database. See LD for complete error Tue Sep 20 2005 08:36:512005 Error:230015013 Host: ls16 Process:scanmgr File: cal_db_file_io.c Line: 226 Function: CAL Processing : SBC Processing Failure opening a file for reading in the Cal Database. File:usr/g/cal_db/kv120/large_sfov/16x0625/large_spot/dasgain5/cal.1.sin Error: 230015035 Cal_db failure See LD for complete error Tue Oct 18 07:20:43 2005 Error: 230015035 Host: ctls3 Process: scanmgr File: cal_db_file_io.c Line: 762 Cal_db failure, error number = 12, reason: Not enough space Tue Sep 20 08:36:512005 Error: 230015035 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 231 Cal_db failure, error number = 98, reason: Address already in use Error: 230015024 A corrupt file has been found in the cal database See LD for complete error Tue Sep 20 08:36:51 2005 Error: 230015024 Host: ls16 Process: scanmgr File: cal_db_file_io.c Line: 1983 Function: CAL Processing : SBC Processing A corrupt file has been found in the cal database. The corrupt file has been removed from the cal database and saved for later analysis. Corrupt File: /usr/g/cal_db/kv120/large_sfov/16x0625/larg
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 Class C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395177 2395162, 2395163, 2399867 LightSpeed Plus/ Ultra MDAS Apps Software (Disk 1, Disk 2 and Patch A) 308I.2_H3.1M5 FW47-03 LS2002M5. IRIX Only. Class C/M Common w/ LS 2X, 3X, & 4X.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.