Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 245551 FEP requested scan abort See the long description for the full error code Tue Jan 3 16:53:23 2006 Suite: LCT1 Host: SBC0 Proc: dataacq Error: 245551 File: tl_fep_events.m 1.18 Method: event_fep_abort_xr Line: 260 Function: Data Acquisition : SBC Processing Scan: 0/0/0 Type: Xray-off Axial RTS Scan Seq: 58001 FEP requested scan abort

Error: 253801 View Length Error. The size of the views received were incorrect See Long Description for complete error Mon Dec 6 12:12:06 2004 Error: 253801 Suite: RHCT Host: FEP Process: CoordinateMode File: fepin.c 1.61 Method: FEPIN_err_handler Line: 2355 Function: Data Acquisition : Scan Data Acquisition Scan: 16016/2/5 Type: Helical Scan Seq: 176001 FEP DSP 0: View Length Error. The size of the views received were incorrect (HDW status: 0x4006b7) (Data Type: 0x1). View Sequence Number: 6537. Possible causes: Slip Ring buffer terminator boards, SCOM board, or FEP board

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2302509-2 SIGNAL BRUSH BLOCK FOR HSA/CTI SYSTEMS See long description for details Part #: 2302509-2 Part Type: GP CONSUMABLES Part Weight: 0.62 LBR Repairable: No Returnable: No Hazardous Material: N Description: SIGNAL BRUSHBLOCK - REPLACEMENT CT/I FRU: Yes Status: Orderable Comments: FLS - 8-7-03, CONSUMABLE COLLECTION PER ANJANA ON PROMOTION THROUGH JUNE 30, 2004 DO NOT CHANGE PRICE PER SHAWN (MAYND) DO NOT DISCOUNT - PER SHAKUN Note: 2302509 FOR H1 (LIGHTSPEED) SYSTEMS That is, DO NOT mix and match ETC and Schliefring Signal Brush Blocks. Any matched pair of Signal Brush Blocks may be installed with either Power Brush Block. See HSA/CTI Non EMC and EMC Gantry Slipping Comparison in Direction 46-018316 for any updates to this information.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.