GE Healthcare · MRExcite2223849225210622238512223817
Application Software does not complete the boot process with TPS Download failure.
Symptom
Application Software does not complete the boot process with TPS Download failure
Error message
2223849 APS EPROM power-up boot failed. (See attachment for extended error message) 2223849 lx-mr NSP TardisReset.cc 1807 APS EPROM power-up boot failed. 2252106 Failed to determine hardware state for APS board within 20 seconds. Error: 2252106 Host: BARLOW_MR Process: NSP File: diags_utils.c Line: 666 Diagnostic software error. Failed to determine hardware state for APS board within 20 seconds The board may be busy loading vxWorks or may be hung. Also check the serial cable for the board and the serial terminal server. 2252106 Failed to determine hardware state for SCP board within 20 seconds 2223849 AGP EPROM power-up boot failed See Long Description Error: 2223849 Host: rmns Process: NSP File: TardisReset.cc Line: 1781 AGP EPROM power-up boot failed. Communication Timed-out from AGP board. Verify that MCP820 boards are installed and seated properly for APS and AGP. Extended Error : MCP820 boards can be identified by the word MCP820 printed on the board extraction lever. Destination Host Unreachable Error message appears when attempting to ping a destination using the "ping" command in a Cshell. 2223849 SCP EPROM power-up boot failed (See attachment for extended error messag
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 2294300-12 SCP2 Board See Long Description 2294300-12 SCP2 Board. FEMC Parts Comments states as follows: If no stock then 2294300-19 is a suitable substitute. 212002950 February 8, 2010.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.