Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600050031818850152230006001

LightSpeed 3.X (Ultra, Plus, QXi) - Octane

Symptom

System will not stay up in applications and shutsdown. System will not stay up in applications and shutsdown

Error message

Error: 260005003 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Thu Apr 29 06:05:11 2004 Error: 260005003 Host: fitzipct Process: procICE File: iceCpu.cxx Line: 343 ICE CPU did not respond to polling within 180 seconds. Recon subsystem is presumed to be down. Some images may have been suspended. Last ping was at Sun May 30 22:21:46 2032 with seq num 1818850152 iceConsole.log: Initializing System Memory (DRAM)... System Memory: 128MB, ECC Enabled (ECC-Memory Detected) PPC1-Bug

Error: 230006001 Process presumed down: vxmonitor-post Thu Jun 22 14:57:42 2006 Error: 230006001 Host: ls01 Process: procICE File: iceCpu.cxx Line: 360 Process presumed down: vxmonitor-post

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2370825 RIP / Motorola Board, 2400 Series Card - replaces all previous part numbers - Oct 14 2004 See long description and attachment 2370825 RIP / Motorola Board, 2400 Series Card, replaces all previous part numbers - 10/14/2004 Part #: 2370825 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 9.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PPC MVME 2400 ASSY, HAWK3 U3D 256MB FRU: Yes Status: Orderable Comments: *** OBSOLETE PART *** This part has been identified as obsolete and no short-term solution has been identified. FE must identify and order a higher or lower level spare part as required. Contact your support engineer, on-line engineer or Parts ID team should you require support. 212012173 10-2-06 2/3/05 (212001261) New RIP-Bd 2370825 only works with LS2002M5 on LS 2X/3X & HS_QXi. LS 4X needs to run LS2002M5.1. To confirm what software version is running in the console, open a shell and type in check_config Even if the system is running LS2002M5, a LFC will probably be required. To determine if a LFC is required, open a shell and type in: showprods (the pipe symbol) grep Helios_MVME enter] if the prompt returns (Helios_MVME24XX_VxWorks_Product) where XX = any number, then no LFC, run pflash. If the prompt returns something else, LFC is required. GOC2/Linux-Pegasus are compatible with RIP-Bd P/N 2370825, (pflash) will need to be run. The following s/w versions support RIP Board 2370825: HS QX/I (IRIX) = 2401246 (101I.2_HSQXIM5), LS 2X (IRIX) 2395178 (2.4.2_H2.4M5); LS 3X (IRIX) = 2395177 (308I.2_H3.1M5), LS 4X (IRIX) = 5116050 (405I.3_H4.0M5.1), HS QX/I (Linux) = 2401245 (103L.4_HSQXIM4), LS 3X (Linux) = 2400882 (308L.5_H3.3M4), LS 4X (Linux) = 5111370 (405L.8_H4.2M4). Can order software via Coakly. Alejandro Cortez
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.