GE Healthcare · CTHiSpeed

System would not boot with PANIC: KERNEL FAULT reported.

Symptom

System would not boot with PANIC: KERNEL FAULT reported. System would not boot and monitor screens turn blank with PANIC: KERNEL FAULT.

Error message

Memory Parity Error in SIMM Sxx Recoverable memory parity error detected by CPU PANIC: KERNEL FAULT - found in SYSLOG See LD for more info Memory Parity Error in SIMM S10 Recoverable memory parity error detected by CPU PANIC: KERNEL FAULT Defective SIMM memory on CPU board caused Memory Parity Error in the SYSLOG Troubleshooting kaiserct 12# cd /usr/adm kaiserct 13# tail -100 SYSLOG | more Jul 11 12:35:05 3D:kaiserct ppp[308]: + killall -v -USR1 gated Jul 11 12:35:05 3D:kaiserct ppp[308]: + 1>& 2 Jul 11 12:35:05 3D:kaiserct ppp[308]: UX:killall: ERROR: process <gated> not found Jul 11 12:35:06 2E:kaiserct savecore: reboot after panic: <0> PANIC: KERNEL FAULT Jul 11 12:35:06 2E:kaiserct savecore: pb 0: 8 <0x380> code:30 Jul 11 12:35:06 2E:kaiserct savecore: pb 1: Memory Parity Error in SIMM S10 Jul 11 12:35:06 2E:kaiserct savecore: pb 2: Recoverable memory parity error detected by CPU Jul 11 12:35:06 2E:kaiserct savecore: pb 3: Memory Parity Error in SIMM S10 TYPE THE FOLLOWING FOR SUSPECTED “MEMORY ERRORS ONLY” IN THE SYSLOG kaiserct 12# cd /usr/adm kaiserct 11# cat SYSLOG | grep Memory Jul 11 12:12:17 2A:kaiserct unix: Memory Parity Error in SIMM S10 Jul 11 12:13:26 2A:kaiserct un

Applies to

Models
HiSpeed
Product
HiSpeed YMS Series HiSpeed Series DXi, FXi, LXi, ZXI, and NXI Including Single / Twin detector and Glass / Metal Tubes. See Long Description for Details Subsystem/Components DXI (NP) FXI,LXI (NP+) ZXI (NP++ NXI (NP+ Twin NXI (NP++ Twin OC O2 running Unix or upgraded to Linux PC Gantry Alighnment Lights Halogen Laser Laser Laser Laser Mechanics Common Electronics firmware is different twin Det/Das Twin Det/DAS Table IMS no Yes Yes Yes Yes Generator Jedi Jedi NP++ Jedi Jedi NP++ Jedi Tube D3142T D3142T D3152T D3142T D3152T PDU Common Options List for V/R 5.5, 6, 7, 9, and 10 System 3D Power 440ma Max Image Disk 20000 Smart Prep Fast Scam 0.7 sec Performix 530 cooling Smart Recon Fast Scan 0.8 sec Asymmetric Scan SmartView Fast Scan 1.0 sec Cardiac gating DemtaScan Helical 30 Max Smart Addition Navigator Helical 60 Max Respiratory Gating Helical Plus Helical 120 Max Reformat Cpmmect {rp Raw Data 1000 max VolumeRendering2 ImageDosk Remote Tilt CT Colono Pioneer Mod Enhanced Cooling The following Options name changes Option (V5.5x,V6.0x,V7.x,V9.x,V10x) Option (V8.1x and V11.x) 3D >> Volume Analysis2 Navigator >> Navigator2,VolumeRendering2,VolumeAnalysis2 CtPerfusion >> CTPersusion2 Reformat >> Functool Options V8.1x and V11.x VolumeAnalysis2 Power 440ma ImageDisk 2000 Helical Plus Smart Prep Fast Scan 0.7 sec Performix 530 cooling Connect Pro Smart Recon Fast Scan 0.8 sec Asymmetric Scan ImageDisk SmartView Fast Scan 1.0 sec Cardiac Gating Pioneer MOD* DentaScan Helical 30 Smart Addition Power 200 ma max Navigator2/VolumeRendering2/ VolumeAnalysis Helical 60 Respiratory Gating Power 250 ma Max Power 300 ma Max Helical 120 Thickness 0.5mm CTColon Power 350 ma Max Raw data 1000 CtPerfusion2 AdvVesselAnalysis Remote Tilt Extended FOV Functool Enhance Cooling Performix Tube VolumeRendering2
Component
2351986 MEMORY256MB See LD for more info 2351986 replaces older 64 MB DIMM 2200771 and 128 MB DIMM 2235085 for YMS O2 cpu Part #: 2351986 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: P9231VR O2MEMORY256MB FRU: Yes Comments: FLS - P9231VR O2MEMORY256MB. ADDEDD PER ASIA SPCN CT-2318NP. ASIA SOURCED (GEYMS). MS20544 ON 2/20/03.
Area
CT