Scanner hardware error
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183908 Axial velocity ACCEL tracking fault See the Long Description for the full error code Mon Aug 27 09:12:45 2007 Error: 183908 Host: STC Process: Axial File: ax_error.c Line: 1114 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 14756/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 09:12:45:860 Log Series: 363 Axial velocity ACCEL tracking fault. current: commanded = 39990 mA current: D/A measured = 39844 mA PDU measured = 39453 mA (HW/SW Error: Retry Rx to verify probable Axial Driver or. Axial Encoder hardware problem.) Error: 183921 Axial conditions at FAULT: See Long Description for complete error Mon Oct 17 08:26:48 2005 Error: 183921 Suite: SMCT Host: STC Process: Axial File: ax_error.c Method: No Method Line: 1114 Function: AXIAL CONTROL : Axial Servo Control Loop Scan: 28848/2/1 Type: Helical Exception Level: Pri/Most Severe Time: 08:26:47:775 Log Series: 141 Axial conditions at FAULT: velocity_fb = 2662 position_fb = 50389 Formatted data: 8100 100 8 7 1 0 1 1 101 1 0 2662 -12655275 13081394 2741 2582 134 133 0x019e38
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2102012 Gantry axial motor brush and cap kit for Lightspeed 1x, CTi, HSA, HLA, and 9800 CT systems. See Long Description for details Part #: 2102012 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: FOR 9800 HLA and HSA GANTRY AXIAL MTR FRU: Yes Status: Orderable Comments: FLS - MOTOR BRUSH & BRUSH CAPS KIT. PEERLESS-WINSMITH REPLACED AS SUPPLIER BY B.D.I. 8-31-01 PERLISA GILLIAN. MS7093 KIT CONTAINS 24 PIECE PARTS, CONSISTING OF: 46-194465P22 VN#819-23-0711-0 (8 PIECES PER KIT) FRU N 46-194465P23 VN#820-07-0200-0 (8 PIECES PER KIT) 46-194465P27 VN#820-07-1300-0 (8 PIECES PER KIT) This kit includes 8 brushes and caps for gantry axial motors used in Lightspeed 1x, CTi, and HSA CT systems as well as 8 brushes and caps for gantry axial motors used in HLA and 9800 CT systems.
- Area
- CT
Full cause & step-by-step resolution