GE Healthcare · CTHiSpeed200306016126152
Beam Obstructed Error while running cals and washed out images
Symptom
Beam Obstructed Error while running cals and washed out images Beam Obstructed Error 126152 and Error: 200306016 when running cals or fastcals and washed out images or artifacts
Error message
Error: 200306016 Received Cal done with status CAL_STATUS_OBSTRUCTED See Long Description for complete error. Mon Feb 5 17:45:27 2007 Error: 200306016 Host: scct01 Process: scanUtilsRx File: detailedCalCommonFuncs.c Line: 1085 Received Cal done with status CAL_STATUS_OBSTRUCTED Error: 126152 Beam-obstruction error See Long Description for complete error. Mon Feb 5 17:45:27 2007 Error: 126152 Suite: CT01 Host: SBC0 Process: calMain File: AirCalJob.m 1.28 Method: jobDone Line: 785 Aircal processing error: Beam-obstruction error. Major Function: Cal Processing; Minor Function: Cal Data Processing.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1 6.1.19.1 for HiSpeed CT/i with with CT/I5 CONSOLE (Octane) - showprods returns Rhapsode Product 6.1.19.1
- Area
- CT
Full cause & step-by-step resolution