Will not Scan with 2254207 SRFD2 RF Amplifier Device Fault 44 RF FET device is shorted Due to the SRFD2 Amplifier.

Symptom

Will Not Scan With RF Errors Will not scan with RF errors

Error message

2254207 SRFD2 RF Amplifier Device Fault 44. See Long Description 2254207 RF FET device is shorted. SRFD2 will attempt to bypass this failure and run in low power mode. Replace Amplifier.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR