Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260118144 (25-0532)LVPS400:heater0 + 160V too low. See Long Description for complete error. Mon Oct 23 08:02:17 2006 Error: 260118144 Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Mon Oct 23 08:02:16 2006 Host : Jedi Ermes # : 260118144 Exception Class : Secondary Severity : Sec/Soft File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Helical Scan: 8818/2/1 (25-0532)LVPS400:heater0 + 160V too low. ^C Z6,Z7= 82h,0h Error: 260118119 (20-0254)Heater Inverter overcurrent(inverter1) See Long Description for full error Sat Apr 15 11:57:29 2006 260118119 orp orp subsys gen exec.c 2150 Sat Apr 15 11:57:27 2006 Host : Jedi Ermes _ : 260118119 Exception Class : Soft Severity : Pri/Soft File : gen exec.c Line_ : 2150 Function : No System Function Reported Scan Type : Helical Scan: 4436/2/1 (20-0254)Heater Inverter overcurrent(inverter1)( 3times). Z6,Z7= 95h,0h
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 2324118-3 VCT-JH4 Auxiliary Equipped. Aux Box See Long Description for details Part#:2324118-3 Part Type: GP CONSUMABLES Part Weight: 2.20 LBR Repairable: No Returnable: No Hazardous Material: N Description: VCT-JH4 AUXILIARY EQUIPED FRU: Yes Status: Orderable Comments: Set as Critical Part for Product CT_VCT, ms04278, 05Oct28 Added per VCT RSPL, ms13452 041205 Picture below
- Area
- CT
Full cause & step-by-step resolution