GE Healthcare · MRSigna22245894062796
High Transmit Gain Is Due To Large Patient Over 250lbs.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2224589 TG reached the high limit. See Long Description Suite: MR01 Host: OC0 Proc: MSG_HANDLR0 Error: 2224589 File: PMH.m 1.29 Method: ps_send_ta Line: 50 TG reached the high limit. Check patient positioning and verify the patient is not touching the transmit coil. Try Auto Prescan again or use Manual Prescan. Note: On later systems, this same error code has been slightly reworded to read as follows: Thu Jun 12 11:06:12 2008 Error: 2224589 Host: MENEMR Process: NSP File: TPS:aps_exec Line: 4062796 TG reached the high limit. TG has been set to 190. Check patient positioning and if desired try Auto Prescan again or use Manual Prescan.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.