Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed230023045230017821470544300000230017801

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

Collimator calibration drift check has failed Cal will continue with the next step - FPA Check.

Error message

Error: 230023045 The Collimator calibration drift check has failed See the long description of the full error code Thu Aug 26 02:56:40 2004 Error: 230023045 Host: ct01 Process: dailyPrepRx File: DailyPrepUtils.c Line: 3155 The Collimator calibration drift check has failed. Full collimator calibration of all stations must be done. Cal will continue with the next step - FPA Check.

Error: 230017821 ACCS CCal: Drift check: The polynomial value has drifted by 0.470544 See the LD for the complete error code Error: 230017821 ACCS CCal: Drift check: The polynomial value has drifted Wed Aug 9 08:18:38 2006 Error: 230017821 Host: qxi1_sbc Process: accs File: collSweepProcessing.c Line: 1635 ACCS CCal: Drift check: The polynomial value has drifted by 0.470544 which is more than tolerance 0.300000

Error: 230017801 ACCS CCal: Sweep Processing failed. Status = -1 See the Long Description for the full error code Tue Nov 30 02:33:04 2004 Error: 230017801 Host: k2_ct2_ls Process: accs File: ccal_prep.c Line: 661 ACCS CCal: Sweep Processing failed. Status = -1

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
06MW29.7 - supports HS QX/I, LS 1X, 2X, 3X, 4X, 5X, BrightSpeed & 7X GRE host consoles, except LS 5X RT16/Xtra See Long Description for more information 06MW29.7 - supports HS QX/I, LS 1X, 2X, 3X, 4X, 5X, BrightSpeed & 7X GRE host consoles, except LS 5X RT16/Xtra CLASSM & CLASSC 5181129 = Class M for 06MW29.7 5192235 = Class C for 06MW29.7 Throw away 06MW29.7 Class C CD, p/n 5182628 ! ! ! replace it with Class C CD, p/n 5192235 REMOTE QUERY OF IG, VIG, DARC2, & VDARC TO DETERMINE IF “WESTVILLE” OR “JARRELL” VERSION · “Westville” computer in IG, VIG, DARC2, & VDARC end of life · IG, VIG, DARC2, VDARC subassemblies use the same computer, whether “Westville” or “Jarrell” · The current Intel “Westville” computer is being replaced by Intel’s “Jarrell” · “Jarrell” (IG, VIG, DARC2, VDARC) targeted for mfg cut-in in Milwaukee & Hino FW39-40’06; Beijing later · “dmidecode” command may be used remotely to determine which is installed in the host console 1. Login to IG, VIG, DARC2, or VDARC as “root” 2. Type dmidecode | less 3. View “Version” line towards top of response If SWV25 , then WESTVILLE installed If SE7520 JR 23 , then JARRELL installed NOTE: In multi-IG configuration where both Westville & Jarrell are installed in same host console (mixed config), the Jarrell IG must be in Slot #1 to optimize speed/performance. WESTVILLE RESPONSE: (all subassemblies use the same Intel Westville computer) $ su - $ dmidecode | less # dmidecode 2.6 # dmidecode 2.2 SMBIOS 2.3 present. 73 structures occupying 2782 bytes. Table at 0x000F1260. Handle 0x0000 DMI type 0, 20 bytes. BIOS Information Vendor: Intel Corporation Version: SWV25 .86B.0257.P42.0504011224 Release Date: 04/01/2005 …etc…. JARRELL RESPONSE: (all subassemblies use the same Intel Jarrell computer) $ su - $ dmidecode | less # dmidecode 2.6 SMBIOS 2.3 present. 59 structures occupying 2536 bytes. Table at 0x000FD4A0. Handle 0x0000 DMI type 0, 20 bytes. BIOS Information Vendor: Intel Corporation Version: SE7520 JR 23 .86B.P.10.00.0087.120820051348 Release Date: 12/08/2005 …etc…. “Jarrell” SOFTWARE RELEASE 06MW29.7 CLASS C AND M TAB #182-DISTRIBUTED 19SEP06 “Jarrell” software release 06MW29.7 Class C & M CDs TAB #182-distributed 19-Sep-06 5181129 = Class M for 06MW29.7 5182628 = Class C for 06MW29.7 - superceded by 06MW29.7 Class C, p/n 5192235 throw away 06MW29.7 Class C CD, p/n 5182628 replace it with Class C CD, p/n 5192235 -> will be TAB #182-distributed with cover letter Class C p/n 5182628, GRE diagnostics sw, incorrectly reads VIG/VRAC & VDARC/VDIP identification data It sees a mismatch between the “expected” and “actual” ID string read from the device The word “Class” is inserted in the “actual” ID string Given this mismatch, it reports a hardware FAILURE when hardware is “good” This issue is fixed on Class C, p/n 5192235 The following steps may be used to determine if the correct Class C software (p/n 5192235) is loaded: 1. Type cd /usr/g/bin 2. Type sum v*test.py 3. Check that the following output: 19976 23 vdarc_test.py 12195 19 vrac_test.py If this output is viewed, then Class C, p/n 5192235, is correctly loaded If this output is not viewed, then uninstall the current Class C software & load Class C, p/n 5192235 5181591 = 06MW29.7 applications software collector 06MW29.7 software release resolves IG, VIG, DARC2, VDARC computer motherboard End Of Life issue. 06MW29.7 software release supports HS QX/I, LS 1X, 2X, 3X, 4X, 5X, & 7X GRE host consoles, except LS 5X RT16/Xtra
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required