The System has the Performed Procedure Step purchased option installed, but Broker is not sending required DICOM fields…

The System has the Performed Procedure Step purchased option installed, but Broker is not sending required DICOM fields for PPS.

Symptom

ConnectPro HisRis Worklist does not Update or update to wrong date ConnectPro HisRis Worklist does not Update or update to wrong date.

Error message

34 Required Field Requested Procedure ID 0040 1001 is zero length in the Worklist response Error seen in the /usr/g/hisris/wlsErrors file: 34 Required Field, (Requested Procedure ID) (0040,1001) is zero length in the Worklist response

34 Required Field Scheduled Procedure Step ID 0040 0009 is zero length in the Worklist response Error seen in the /usr/g/hisris/wlsErrors file: 34 Required Field, (Scheduled Procedure Step ID) (0040,0009) is zero length in the Worklist response

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution