Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Error: 260001012 The Rotating Controller Interface Bus (RCIB) fault line has been asserted See Long Description for the full error code Sat Jun 9 14:37:06 2001 Error: 260001012 Host: OBC Process: HOP File: hop_fault_state.c Line: 137 Function: No System Function Reported Scan: 9553/3/1 Type: Axial Exception Level: Pri/Most Severe Time: 14:37:06:030 Log Series: 366 The Rotating Controller Interface Bus (RCIB) fault line has been asserted. None of the sub-systems have reported a problem. Check nodes or fault line for failure.

Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the CCB See the long description for the full error code Tue Nov 13 21:06:38 2007 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 2552/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:06:38:745 Log Series: 937 CAN serial link error. An CC error was detected when sending a message to the CCB. Message code: 1004 Node ID: 52 Object ID: 0 Send status: 5

Do Not Use - use one of the other Error: 260001008 The OBC failed to connect with the ? (DCB and/or the CCB) sub-system in the allotted time Open Long 

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-170021P81 20A M-DAS power supply fuse See Long Description for details Part #: 46-170021P81 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: LITTELFUSE 323020, 20A,250V, 3AB FRU: Yes Status: Orderable Comments: FLS - 20A,250V
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required