Pular para o conteúdo principal
GE Healthcare · CTLightSpeed10032

63820 LightSpeed 4.X with GRE and M-DAS16 Gantry

Symptom

Customer is unable to archive using 5.2GB or 9.1 GB MOD media

Error message

Error: 10032 Media is not formatted or of unknown format See Long Description for the full error Sun Jun 19 22:29:28 2005 Error: 10032 Host: ct01 Process: arsjob File: ArchiveDevice.c Line: 4677 Media is not formatted or of unknown format

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
05MW14.5 for all systems with GRE/Xtream consoles on HiSpeed QX/i and LightSpeed 1.X, 2.X, 3.X, 4.X, 5.X Series(QXi, PLUS, ULTRA, LS16, PRO 16, RT). Obsoleted by FMI 25365 deploy FMI 25365 It is the Service person's responsibility to use the appropriate documentation when performing and verifying service. Refer to FMI 25365: 06MW03.5 SW Update for CT-LS 1X/2X/3X/4 DIR 5161945-100, which can be found on the Distributed Technical CDROM/DVD, or online Links to any needed documentation is found in the Product concepts as an attachment called "ProductDocumentLinks"
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues