GE Healthcare · MRSigna2225643

Erbtec Faults Due to Door Knob Capacitor Arcing In RF Amplifier.

Symptom

Unable to scan with Erbtec errors

Error message

2225643 ERBTEC RF Amplifier fault 57 +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay 2225643 ERBTEC RF Amplifier fault 57. +3kV power supply unacceptable within 0.5 to 6 seconds after closing HV start relay. Possible defective HV rectifier/filter board, bad relay in AC Switching Module, or bad A/D on Processor Board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
46-305000P57 Erbtec Doorknob Capacitor
Area
MR

Full cause & step-by-step resolution