LightSpeed 5.X (60kW)
Symptom
Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .
Error message
Error: 260116313 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out. see LD Tue May 16 22:03:29 2006 Error: 260116313 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 349 Running: System Reset Tue May 16 21:37:53 2006 Host : Table Gantry Processor Ermes # : 260116313 Exception Class : Safe State Severity : Secondary File : subsys_mgr.c Line# : 349 Function : SYSTEM COMMUNICATIONS : Control Network Scan Type : None/Unknown Scan: 0/0/0 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out. Error: 244642 Fwmgr sensed firmware watchdog timeout. See LD for complete error Error: 244642 Fwmgr sensed firmware watchdog timeout Fri Aug 5 14:37:54 2005 Error: 244642 Host: ct01 Process: fwmgr File: fwmgr_events.c Line: 8925 Running: System Reset Fwmgr sensed firmware watchdog timeout. Error: 244629 Lost Communication with TGP(LAN Watchdog) See Long Description for the full error code Tue Nov 22 08:12:44 2005 Error: 244629 Host: law-ct1 Process: fwmgr File: fwmgr_fw_if.cxx Line: 404 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Lost Communication with TGP(LAN Watchdog)
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 See long description for details 5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 Part #: 5126615 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASTEC Digital PS Assembly FRU: Yes Status: Orderable Comments: Added per RT / GDAS RSPL
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required