Pular para o conteúdo principal
GE Healthcare · CTLightSpeed260125041900824897741749120260134703903212

LightSpeed 5.X (60kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260125041 Motion without command in Elevation drive See LD for more details Wed Jun 14 12:48:32 2006 Error: 260125041 Host: tgp Process: tgp_subsys File: ta_callback_gt.c Line: 508 Wed Jun 14 12:48:31 2006 Host : Table Gantry Processor Ermes # : 260125041 Exception Class : Abort Severity : Primary File : ta_callback_gt.c Line# : 508 Function : No System Function Reported Scan Type : Helical Scan: 412/4/1 Motion without command in Elevation drive Current pos = 900824 Last stop pos = 897741 Tolerance = 3000 Time since last stop = 749120 Last target pos = 1 Hardware status = 0

Error: 260134703 Elevation moved without command. See LD for more details Wed Jun 14 12:48:34 2006 Error: 260134703 Host: tgp Process: tgp_subsys File: gtcb_emc_movechk.cxx Line: 530 Wed Jun 14 12:48:34 2006 Host : Table Sub-System Ermes # : 260134703 Exception Class : Secondary Severity : Pri/Soft File : gtcb_emc_movechk.cxx Line# : 530 Function : No System Function Reported Scan Type : None/Unknown Scan: 412 Elevation moved without command. If characterize completed appears above, this isnt error. If this appears many times, check potentiometer. Current position = 903212 um Last stop position = 897741 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2337217 elevation encoder assembly used on HPower Systems. See Long Description for details Part #: 2337217 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ENCODER ASSY, ELEVATION - H POWER FRU: Yes Status: Orderable Parts Offered as: Current: 2337217 Harvested: Not Available Exchange: Not Available Comments: FLS - ENCODER ASSY, ELEVATION - H POWER. ADDED PER CT PCN 220839. MS20544 3/31/03.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues