Pular para o conteúdo principal
GE Healthcare · CTHiSpeed182209

63199 HiSpeed CT/i 4.X with SGI Indigo Console and HSA Gantry

Symptom

Cradle will move a few inches and stop. Cradle will move in a few inches and stop

Error message

Error: 182209 HW error: No Encoder C pulse within 410 mm (two turns of the Potentiometer). See Long Description for full error Sun Nov 6 12:33:30 2005 Error: 182209 Host: ETC Process: Cradle File: cmcCradlePos.c 1.21 2/5/96 Line: 84 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 7462/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:33:30:210 Log Series: 58 HW error: No Encoder C pulse within 410 mm (two turns of the Potentiometer).

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues