Intermittent loss of signal due to interference with the system side connector.
Symptom
Intermittent loss of signal with 8 channel Body or CTL Array coils See Long Description Subject: MR SMILe: FMI 60686 and 60687, SIGNA 8CH CTL AND BODY ARRAY SYSTEM CONNECTORS Effectivity: All 1.5T Signa systems with 8 channel Body or CTL array coils (p/n 2377425-2 and 2377426-2, respectively). This FMI only applies to coils with serial numbers less than 750. All coils with serial numbers greater than 750 shipped with all current changes. Description: The original design of the system connector housings on these coils can cause intermittent loss of signal due to possible interference with the system side connector. The most common symptoms of this issue include customer complaints of intermittent image quality issues without driver faults when using these 8ch coils. A new connector has been designed to minimize these issues and improve the overall fit of the 8ch connector. SMILe 1/27/2005
Error message
2224725 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan) See Long Description Error: 2224725 Host: hopgemr1a Process: NSP File: APS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan).
Applies to
- Models
- Excite, Signa
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.