Infinia Gantry GReset fails first thing in the morning with Error #155 on Axis 8 Linear, also on Axis 5 or 6 S…

Infinia Gantry GReset fails first thing in the morning with Error #155 on Axis 8 Linear, also on Axis 5 or 6 Swivel.

Symptom

Infinia Gantry GReset fails first thing in the morning with Error #155 on Axis 8 Linear, also on Axis 5 or 6 Swivel. Sometimes after multiple GReset, Gantry will Initialize all Axis Infinia Gantry GReset fails first thing in the morning with Error #155 on Axis 8 Linear, also on Axis 5 or 6 Swivel. Sometimes after multiple GReset, Gantry will Initialize all Axis. Error # 155 TEST: ANALOG SWITCH FAILURE, AXIS=LIN ERROR_HVG_DETAILED:.: Gantry error 155 .:.:Gantry error 155.:.: Detailed HVG error information:009B 0008 Error # 155 TEST: ANALOG SWITCH FAILURE, AXIS=SW1 ERROR_HVG_DETAILED:.: Gantry error 155 .:.:Gantry error 155.:.: Detailed HVG error information:009B 0005 Contributor 212011798 Steele February 16 2007

Error message

Error #155 (Error 155) TEST: ANALOG SWITCH FAILURE, AXIS=$ Infinia gantry Error #155 comes as part of the motion Safety tests. The Analog switch is a switch in the SEM. This error could come from: A faulty SEM ( where this switch is defective ..) or From some off-set on the SEM to Elmo driver - command signal, or Some off-set in the ELMO Driver

Applies to

Models
Infinia
Product
Infinia II Infinia II Acquisition System
Component
*Gantry Gantry Gantry
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.