Pular para o conteúdo principal
GE Healthcare · CTLightSpeed230015023

LightSpeed 4.X - Octane

Symptom

Message appears on the scan monitor for 3 seconds during selection of new patient "None of the KV stations are fully calibrated. Calibration is required prior to scanning

Error message

Error: 230015023 Failure loading zslope fields in cal module header Wed Aug 25 16:54:09 2004 Error: 230015023 Host: racconri20 Process: scanmgr File: cal db funcs.c Line: 388 Function: CAL Processing : SBC Processing Failure loading zslope fields in cal module header. File: usr/g/cal db/kv80/large sfov/4x250/small spot/zscal.1

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
405I.2_H4.0M5 for LightSpeed 4.X (LightSpeed 16) with H16 Console (GOC1)(2341923, 2341923-2) - showprods returns 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395173 1204I.2_ H4.0M5 CLASSM LightSpeed/ HiSpeed QX/I Class M Software 2395176 2395170, 2395171, 2399867 LightSpeed 16 Applications Software (Disk 1, Disk 2 and Patch A) 405I.2_H4.0M5 FW47-03 LS2002M5. IRIX only. Class C/M Common w/ LS 2X, 3X, & 4X.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required