P500D terminates digital record exposure terminated below 10 mS.

Precision 500D Short exposures during Digital record or during mR/mAS calibration on Fluoro Tube

Symptom

P500D terminates digital record exposure terminated below 10 mS. When attempting to recalibrate mR/mAS on fluoro tube, actual exposure time is 2 mS instead of expected 100 mS displayed before exposure. Overheat tube exposures times are normal on RAD overhead X-Ray tube. Tube warmup can be completed normally on Fluoro tube. mR/mAS can be completed normallly on the overhead X-Ray tube.