GE Healthcare · CTLightSpeedVCT260107204
LightSpeed 5.X (60kW)
Symptom
The positioning lights will not come on when the Patient was being positioned for a scan The laser lights will not come on when the Patient was being positioned for a scan
Error message
Error: 260107204 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry. See Long Description for full error Wed Apr 7 21:18:13 2004 Error: 260107204 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 230 Wed Apr 7 21:18:13 2004 Host : Table Gantry Processor Ermes # : 260107204 Exception Class : Abort Severity : Secondary File : amc_mgr.c Line# : 230 Function : AXIAL CONTROL : Axial Communications Scan Type : None/Unknown Scan: 0/0/0 Axial Motor Control (AMC) logic could not configure the AXIAL drive via the CAN network during configuration. Reset scan hardware to retry.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P15 (250V,8A) Fuse used in axial dynamic brake circuit as CT2A1A10F1 and found in FAN ASSY- GANTRY, and F4, F5 on HEMRC Interface Board and other places See long description for details Part #: 46-170021P15 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: 008.000A 0250V 3AB SLOW UL/ FRU: Yes Status: Orderable Comments: FLS - FUSE 8 AMP - 250 VOLT 3AB
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.