Pular para o conteúdo principal
GE Healthcare · MRExciteOvation22185922218630221850122481302248122

CWA ID62427 The Download Of A Prescribed Series Failed.

Symptom

The Download Of A Prescribed Series Failed

Error message

2218592 The MrMail send of real time filter to TPS timed out Thu Jun 3 05:32:17 2004 2218592 qstmr1 scn msaUtil.m 175 The MrMail send of real time filter to TPS timed out.

2218630 The download of a prescribed series failed. 2218630 Host: Andys3T Process: scn File: RxManagerIF.m Line: 556 The download of a prescribed series failed. Internal status code is 2218501.

2248130 APS DC Setup DRF Channels Error. DRFs can only support -x channels 2248130 Host: UMIMR3 Process: NSP File: APS:DacqControl::processPkt Line: 513 APS DC setup DRF channels error DRFs can only support -1 channels Software error, please perform a TPS Reset

2248122 Number of Receiver Prescribed is Over the DRF Capability. 2248122 Host: UMIMR3 Process: NSP File: APS:DrfDriver::setupChannel Line: 20 Number of receiver prescribed is over the DRF capability. Maximum receivers supported at this system is 7 Please change prescription.

Applies to

Models
Excite, Ovation
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required