Pular para o conteúdo principal
GE Healthcare · MRExcite2254253

Unable to Scan Due to 3T MKS RF Amplifier will not come ready. 3T Excite HD 12.0. 2254253 3T RF Amplifier PA Filament un…

Unable to Scan Due to 3T MKS RF Amplifier will not come ready. 3T Excite HD 12.0. 2254253 3T RF Amplifier PA Filament under Voltage fault38

Symptom

Unable to Scan with 3T MKS RF Amplifier will Not go Ready MKS amp powers up but will not come ready.

Error message

2254253 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). 2254253 NSP SCP:../rf_nb_contro 1460 memp 8HRBRAIN protocol:DQA_8ch_Brain/1 3T RF Amplifier PA Filament under Voltage (fault 38). The PA Filament supply output was below specification(15V - 5%). Possible Causes: Defective supply or short on supply output. If the fault persists, replace the RF amplifier

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues