Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26011631950132

LightSpeed 5.X (80/100kW)

Symptom

When pressing Alignment Light Button, Gantry rotates but no lasers come on and the ORP reset indicator flashes on the gantry display

Error message

Error: 260116319 State Response Timeout Occurred See Long Description for full error Fri Sep 2 07:22:58 2005 Error: 260116319 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 2343 Fri Sep 2 07:22:55 2005 Host : Table Gantry Processor Ermes # : 260116319 Exception Class : Secondary Severity : Pri/Soft File : subsys_mgr.c Line# : 2343 Function : No System Function Reported Scan Type : None/Unknown Scan: 50132/1/0 State Response Timeout Occurred Subsystem: Table Gantry Processor State: 12 Non-Responding App Id: 100 Timeout Value: 1000 msec.

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2105617-5 (replace 2105617-4) SINGLE, TERMINATED LASER DIODE ASM - used on H2 and HP gantry See long description for details SINGLE, TERMINATED LASER DIODE ASSEMBLY Part #: 2105617-5 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SINGLE, TERMINATED LASER DIODE AMP CONNECTOR, 105MM LONG, PHOTONIC FRU: Yes Status: Orderable Part #: 2105617-4 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: SINGLE, TERMINATED LASER DIODE AMP CONNECTOR, 105MM LONG FRU: Yes Status: Orderable Comments: FLS - H2 GANTRY. COMMON TO MFG MS26561 DSM 1/7/01 PART OF 2229971
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.