GE Healthcare · CTHiSpeed183340
HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183340 The rotor control board has detected HVDC Bus voltage when none expected. See the Long Description for the complete error code Tue Jul 1 11:53:31 2003 Error: 183340 Host: OBC Process: Generator File: gen_power.c 1.13 Line: 337 Function: SYSTEM MONITORING : HVDC Monitor Scan: 858/5/1 Type: Helical Exception Level: Pri/Soft Time: 11:53:31:975 Log Series: 675 The rotor control board has detected HVDC Bus voltage when none expected. The Gentry I/O indicates the backup pilot relay is de-energized. HVDC Bus voltage: 46V Expected: 0V Limit: 24V
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2139289-2 CPDU Control Bd. Replaces part 2139289 Used In CPDU - 2269902. 2139289-2 CPDU Control Bd. for Compact PDU Part #: 2139289-2 Part Type: GP REPAIRABLES Part Weight: 5.35 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: 2139289-2 : PDU CONTROL BOARD W/ REW FRU: Yes Status: Orderable Comments: FLS - PDU CONTROL BOARD REWORK ASSEMBLY. THIS P/N 2139289-2, IS P/N T3202KF (2139289) WITH A REWORKED SUB-ASSEMBLY. ADDED PER PCN 220525 BY MS20544 ON 7/17/02. 9/10/02- 2139289-2 REPLACES 2139289 PER PCN 220525. MS12464 11/08/04 SSO# 212001261 FE's are to return defectives and unused/consigned back via FED Ex over night Item #16 A6 panel is the control board
- Area
- CT
Full cause & step-by-step resolution