Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26000104950133244586

LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2

Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Error: 260001049 Communications with the CCB sub-system has not been established. Scanning aborted. See Long Description for full error Thu Mar 31 07:41:47 2005 Error: 260001049 Host: OBC Process: HOP File: hop_misc.c Line: 265 Function: No System Function Reported Scan: 50133/1/1 Type: Axial Exception Level: Secondary Time: 07:41:47:000 Log Series: 111 Communications with the CCB sub-system has not been established. Scanning aborted

Error: 244586 the following hardware subsystems reported a failure when initializing: OBC SS (Coll/Das) See the Long Description for the complete error code Error: 244586 Host: qx-ct Process: fwmgr File: HwSetUp.c Line: 2235 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The following hardware subsystems reported a failure when initializing: OBC SS(Coll/Das)

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.