GE Healthcare · CTLightSpeed2600010495013326000100726000151450004260001022
Scan aborts with CCB communications error
Symptom
Scan aborts WITH Communications with the CCB sub-system has not been established. Scan aborts WITH Communications with the CCB sub-system has not been established
Error message
Error: 260001049 Communications with the CCB sub-system has not been established. Scanning aborted. See Long Description for full error Thu Mar 31 07:41:47 2005 Error: 260001049 Host: OBC Process: HOP File: hop_misc.c Line: 265 Function: No System Function Reported Scan: 50133/1/1 Type: Axial Exception Level: Secondary Time: 07:41:47:000 Log Series: 111 Communications with the CCB sub-system has not been established. Scanning aborted Error: 260001007 CAN serial link error. A CC error was detected when sending a message to the CCB See the long description for the full error code Tue Nov 13 21:06:38 2007 Error: 260001007 Host: OBC Process: HOP File: hop_cc_send.c Line: 196 Function: No System Function Reported Scan: 2552/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 21:06:38:745 Log Series: 937 CAN serial link error. An CC error was detected when sending a message to the CCB. Message code: 1004 Node ID: 52 Object ID: 0 Send status: 5 Error: 260001514 RCIB Bus: Fault Line Interrupt Occurred. Timeout Waiting for Host Response to Fault Line Verification. See long description for full error Wed Apr 27 01:00:51 2005 Error: 260001514 Host: orp Process: orp_subsys File: notiftsk.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT