GE Healthcare · CTLightSpeed18221120525750009085218222633542
LightSpeed 3.X (Ultra, Plus, QXi) - GOC2
Symptom
Grinding noise coming from the table while driving cradle in or out. Table may move slower than normal. Customer may hear a grinding noise coming from the table while driving cradle in or out. Table may move slower than normal.
Error message
Error: 182211 Velocity error detected by hardware (LM628 excessive position error). See the Long Description for the full error Tue Aug 29 08:09:30 2006 Error: 182211 Host: ETC Process: Cradle File: cmcLm628IsrTsk.c 1.19 10/16/9 Line: 185 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 20525/2/1 Type: Helical Exception Level: Pri/Soft Time: 08:09:30:260 Log Series: 595 HW error: Velocity error detected by hardware (LM628 excessive position error). Expected Velocity 75000 (.001 mm/sec) Actual Velocity 0 (.01 mm/sec) Pos Err Tol 2540 (.01 mm) Cur Pos 90852 (.01 mm) Cradle Dir: 1 (0 = Out 1 = In [Gantry]). Error: 182226 Warning: Cradle did NOT reach Scan Position. see Long Description for complete error Fri Sep 1 15:04:38 2006 Error: 182226 Host: ETC Process: Cradle File: cmcShutDown.c 1.54 2/5/96 Line: 184 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 1145/2/5 Type: Axial Exception Level: Pri/Soft Time: 15:04:38:470 Log Series: 449 Warning: Cradle did NOT reach Scan Position. Start Position 33542 (.01 mm) Scan/End Position 32044 (.01 mm) Distance (Start-Scan) 1498 (.01 mm) Current Position 25299 (.01 mm) Tolerance 25 (.01 mm)
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GOC2 This is HP CPU host running LINUX with Pegasus IG. 308L.8_H3.3M4 ~ 4 or 8 slice MDAS. ~ Could be an Ultra, Plus or QXi system. Must run swokinstall -p to verify. ~ See Long description for details . LightSpeed 3.X MDAS4 Gantry and GOC2 Gantry - 2281177-2, DAS - 2272768, Console – 2341104-2, Table - 2276776, PDU - 2269902 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.